Thông Số Kỹ thuật
|
Accuracy specifications
|
Function
|
Range
|
Resolution
|
Accuracy ±
([% of Reading] + [Counts])
|
DC Millivolts
|
600.0 mV
|
0.1 mV
|
0.5 % + 2
|
DC Volts
|
6.000 V
60.00 V
600.0 V
|
0.001 V
0.01 V
0.1 V
|
0.5 % + 2
|
Auto-V LoZ true-rms
|
600.0 V
|
0.1 V
|
DC, 45 Hz to 500 Hz: 2.0% + 3
500 Hz to 1 kHz: 4.0% + 3
|
AC millivolts true-rms
|
600.0 mV
|
0.1 mV
|
45 Hz to 500 Hz: 1.0% + 3
500 Hz to 1 kHz: 2.0% + 3
|
AC volts true-rms
|
6.000 V
60.00 V
600.0 V
|
0.001 V
0.01 V
0.1 V
|
45 Hz to 500 Hz: 1.0% + 3
500 Hz to 1 kHz: 2.0% + 3
|
Continuity
|
600 Ω
|
1 Ω
|
Beeper on < 20 W, off > 250 W;
detects opens or shorts of 500 µs or longer
|
Ohms
|
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
|
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
|
0.9% + 2
0.9% + 1
0.9% + 1
0.9% + 1
0.9% + 1
1.5% + 2
|
Input characteristics
|
Function
|
Input impedance
(nominal)
|
Common mode rejection ratio
(1 kW unbalanced)
|
Normal mode rejection
|
Volt AC
|
> 5 MΩ < 100 pF
|
>60dB at dc, 50 or 60Hz
|
-
|
Volts DC
|
> 10 MΩ < 100 pF
|
>100dB at dc, 50 or 60Hz
|
> 60 dB at 50 or 60Hz
|
Auto-V LoZ
|
~3 kΩ < 500 pF
|
> 60dB at dc, 50 or 60Hz
|
-
|
Ohms
|
Open circuit test voltage
|
Full scale voltage
|
Short circuit current
|
To 6.0 MΩ
|
To 6.0 MΩ
|
< 2.7 V dc
|
< 0.7 V dc
|
< 0.9 V dc
|
< 350 µA
|
Diode Test
|
< 2.7 V dc
|
2.000 V dc
|
< 1.2 mA
|
General specifications
|
Function
|
Description
|
Maximum voltage between any terminal and earth ground
|
600 V
|
Surge protection
|
6 kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2
|
W Fuse for A input
|
11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)
|
Display
|
Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec
|
Temperature
|
Operating: -10°C to +50°C; Storage: -40°C to +60°C
|
Humidity
|
0% to 90% to 35°C
75% to 40°C
45% to 50°C
|
Temperature coefficient
|
0.1 x (specified accuracy/°C) (< 18°C or > 28°C)
|
Operating altitude
|
2,000 meters
|
Battery
|
9 Volt Alkaline, NEDA 1604A/IEC 6LR61
|
Battery life
|
Alkaline: 400 hours typical, without backlight
|
Safety compliances
|
ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
|
Certifications
|
UL, CSA, TUV, N10140 , VDE
|
IP rating
(dust & water protection)
|
IP42
|