Thiết bị kiểm tra dòng rò Hioki ST5541
Thông số kỹ thuật
■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement methods |
Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground. |
Measurement modes |
Leak current measurement, voltage measurement, safety conductor current measurement |
Standards compliance |
Electrical Appliances and Materials Safety Act |
Leak current measurement |
Ground leak current, 3 types of contact current, free current measurement, 3 types of enclosure leak current |
Measurement current |
DC, AC (true rms, 15 Hz to 1 MHz), AC+DC (true rms, 15 Hz to 1 MHz), AC peak (15 Hz to 1 MHz) |
Measurement ranges |
DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA |
Measurement accuracy (current measurement) |
DC measurement: ±2.0% rdg. ±6 dgt. (typ.) |
Interfaces |
External I/O, USB 1.1 (communications), RS-232C |
Functionality |
Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc. |
Power supply |
100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power |
Target device power supply input |
100 to 240 V AC, 50/60 Hz, Rated current input from terminal block: 20 A |
Target device power supply output |
Output from terminal block: 20 A, Output from outlet: 15 A |
Dimensions and mass |
320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz) |
Accessories |
Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, User guide ×1, CD-ROM ×1 |
■ ST5540, ST5541 List of functions (Network)
Item |
ST5540 |
ST5541 |
Network A (Electrical Appliances and Materials Safety Act) |
○ |
○ |
Network B (Medical-use electrical devices) |
○ |
- |
Network C (IEC 60990) |
○ |
○ |
Network D (UL) |
○ |
○ |
Network E (General-purpose 1) |
○ |
○ |
Network F (General-purpose 2) |
○ |
○ |
Network G (IEC 61010-1) |
○ |
○ |
■ List of functions (Major functions)
Item |
ST5540 |
ST5541 |
Power on polarity switching function |
○ |
○ |
Rated current 20 A |
○ |
○ |
Function for checking for blown fuses |
○ |
○ |
Frequency band switching |
○ |
- |
110% voltage output terminal (T3 terminal) |
○ |
- |
S10, S12, S13, E terminal |
○ |
- |
■ List of functions (Testing leakage current mode)
Item |
ST5540 |
ST5541 |
Earth leakage current |
○ |
○ |
Touch current |
○ |
○ |
Patient auxiliary current |
○ |
- |
Patient leakage current |
○ |
- |
Total patient leakage current |
○ |
- |
Free current |
○ |
○ |
Enclosure - Earth leakage current |
○ |
○ |
Enclosure - Enclosure leakage current |
○ |
○ |
Enclosure - Line leakage current |
○ |
○ |
Patient leakage current I |
○ |
- |
Patient leakage current II |
○ |
- |
Patient leakage current III |
○ |
- |
Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh
Địa chỉ: Số 9,Ngõ 461 Nguyễn Văn Linh, Q. Long Biên, TP.Hà nội
Tel: 024-38751616 và 024-36790555 Hotline: 0904499667 - 0988 764 055
Email: binhminhthuha@gmail.com
============================
Chi nhánh Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh Tại TPHCM
Địa chỉ: Số 1331/15/144 Đường Lê Đức Thọ,Phường 14, Q. Gò Vấp, TP.Hồ Chí Minh
Tel: 024-38751616 và 024-36790555 Hotline: 0988 764 055